Numerical modeling of continuous-flow left ventricular assist device performance

Tely'shev D. V., Petukhov D., Selishchev S.
International Journal of Artificial Organs
Vol.42, Issue11, P. 611-620
Опубликовано: 2019
Тип ресурса: Статья

DOI:10.1177/0391398819852365

Аннотация:
Responses of five rotary blood pumps, namely HeartAssist 5, HeartMate II, HeartWare, Sputnik 1, and Sputnik 2, were extensively assessed in six test cases using a mathematical model of the cardiovascular system. Data for the rotary pumps were derived from pressure–flow curves reported in the literature. The test cases were chosen to attempt to cover most common clinical conditions, such as partial or full support or transitions between different levels of ventricular support. The investigated parameters are collected in a table and presented in figures, such as pressure–volume loops, H-Q curves, pump flow, and aortic pressure waveforms. HeartAssist, Sputnik 1, and Sputnik 2 pumps provide comparable level of aortic pressure, pump flow pulsatility PI(QP), and aortic pressure pulsatility PI(AoP) due to the similarity of pressure–flow characteristic curves of these pumps. HeartMate II provides a minimal backflow among other investigated rotary blood pumps due to the maximum pressure head a
Ключевые слова:
Continuous-flow left ventricular assist device; hemodynamics; preload sensitivity; rotary blood pump; simulation; static H-Q curves
aortic pressure; Article; cardiovascular system; hemodynamic parameters; human; mathematical model; pulsatility index; simulation; adverse device effect; biological model; heart assist device; heart rate; hemodynamics; physiology; pulsatile flow; Heart Rate; Heart-Assist Devices; Hemodynamics; Humans; Models, Cardiovascular; Pulsatile Flow
Язык текста: Английский
ISSN: 1724-6040
Tely'shev D. V. Dmitrij Viktorovich 1984-
Petukhov D.
Selishchev S.
Телышев Д. В. Дмитрий Викторович 1984-
Петюхов Д.
Селишчев С.
Numerical modeling of continuous-flow left ventricular assist device performance
Текст визуальный непосредственный
International Journal of Artificial Organs
Wichtig Editore
Vol.42, Issue11 P. 611-620
2019
Статья
Continuous-flow left ventricular assist device hemodynamics preload sensitivity rotary blood pump simulation static H-Q curves
aortic pressure Article cardiovascular system hemodynamic parameters human mathematical model pulsatility index simulation adverse device effect biological model heart assist device heart rate hemodynamics physiology pulsatile flow Heart Rate Heart-Assist Devices Hemodynamics Humans Models, Cardiovascular Pulsatile Flow
Responses of five rotary blood pumps, namely HeartAssist 5, HeartMate II, HeartWare, Sputnik 1, and Sputnik 2, were extensively assessed in six test cases using a mathematical model of the cardiovascular system. Data for the rotary pumps were derived from pressure–flow curves reported in the literature. The test cases were chosen to attempt to cover most common clinical conditions, such as partial or full support or transitions between different levels of ventricular support. The investigated parameters are collected in a table and presented in figures, such as pressure–volume loops, H-Q curves, pump flow, and aortic pressure waveforms. HeartAssist, Sputnik 1, and Sputnik 2 pumps provide comparable level of aortic pressure, pump flow pulsatility PI(QP), and aortic pressure pulsatility PI(AoP) due to the similarity of pressure–flow characteristic curves of these pumps. HeartMate II provides a minimal backflow among other investigated rotary blood pumps due to the maximum pressure head a